- Antigen information
Access number: P19429
Gene ID: 7137
Genetic symbols: TNNC1, TNNI3
Protein name and synonyms:
troponin I, heart muscle (cardiac troponin I)
Target species: Human
- Test format
Detection range:
0.38 pg / ml – 625 pg / ml
Recommended dilution (serum and plasma): 2 times
Sensitivity: 0.38 pg / ml
Specificity:
This ELISA antibody pair detects human cTnI. Other species not determined.
Number of targets detected: 1
Species detected: Human
Compatible sample types:
Cell culture supernatants Plasma Serum
Design Principle:
Sandwich-based
Detection method: colourimetric
Quantitative / semi-quantitative:
Quantitative
Solid support: 96-well microplate
- Product specifications
Size: 1, 2, or 5 x 96-Well Strip Microplate Kit
- Product characteristics
Additional notes
- Strip plates and additional reagents allow for use in multiple experiments
- Quantitative protein detection
- Set normal range
- Best Products for Antibody Array Data Confirmation
- Application notes
Kit components
- Precoated 96-Well Strip Microplate
- Wash buffer
- Stop solution
- Test diluent (s)
- Lyophilized standard
- Biotinylated detection antibody
- Streptavidin-conjugated HRP
- Single-pass substrate TMB
Other Necessary Materials
- Distilled or deionized water
- Precision pipettes to deliver volumes from 2 µl to 1 µl
- Adjustable 1-25 µl pipettes for reagent preparation
- Graduated cylinders of 100 µl and 1 liter
- Tubes for preparing standard and sample dilutions
- Absorbent paper
- Microplate reader capable of measuring absorbance at 450 nm
- Log-log graph paper or computer and software for ELISA data analysis
Protocol outline
- Prepare all reagents, samples, and standards as directed in the manual.
- Add 100 µl of standard or sample to each well.
- Incubate 2.5 h at RT or O / N at 4 ° C.
- Add 100 µl of prepared biotin antibody to each well.
- Incubate 1 h at RT.
- Add 100 µl of prepared streptavidin solution to each well.
- Incubate 45 min at RT.
- Add 100 µl of TMB One-Step Substrate Reagent to each well.
- Incubate 30 min at RT.
- Add 50 µl of Stop Solution to each well.
- Read at 450 nm immediately.
Storage stability
The complete kit can be stored at -20 ° C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. The kit can be stored at 4 ° C for up to 6 months. For long-term storage, it is recommended to store at -80 ° C.
Use
- This product is provided for LABORATORY RESEARCH USE ONLY.
- It is not for diagnostic or therapeutic use.